An Isocratic Method Development and Validation for simultaneous estimation of Rabeprazole and Mosapride in Tablet Dosage Forms by using RP-HPLC
Abstract
Objective: The present work was undertaken with the aim to develop and validate a rapid and consistent RP-HPLC method in which the peaks will be appear with short period of time as per ICH Guidelines.
Method: The HPLC separation was achieved on an Inertsil-C18 ODS column (250 X 4.6 mm; 5 μ) column in an Isocratic Mode. The mobile phase composed of Methanol [HPLC Grade] (55 %) and Buffer (45 %) [pH 4.0 adjusted with triethylamine]. The flow rate was monitored at 1.0 mL/min. The wavelength was selected for the detection was 276 nm.
Results: The retention times found for rabeprazole and mosapride was 2.946 and 4.186 min respectively. The % recovery was 99.98- 100.03 for rabeprazole and 99.97 - 100.02 for mosapride. The linearity was established in the range of 20-80 μg/mL for both rabeprazole and mosapride. The LOD for rabeprazole and mosapride were 0.01 and 0.035 μg/mL respectively. The LOQ for rabeprazole and mosapride were 0.032and 0.11 μg/mL respectively.
Conclusion: The proposed method was adequate sensitive, reproducible, and specific for the determination of rabeprazole and mosapride in bulk as well as in tablet dosage forms.